It was reported that the arctic sun device had an overshoot.
It was stated that the therapy started overnight, the target temperature was 36c, the current temperature was 33.
9c and the water temperature was 30.
7c.
The nurse stated that the event log showed alert 11 (patient temperature 1 below low patient alert) and alert 045(ac power lost).
The nurse had put the device in manual mode, and the water temperature rose only to 34c.
When the device was placed back to the automatic mode, the water temperature was down to 29c.
The mss suggested changing the devices and to inform the doctor that the patient was under the target.
Per follow up information received via phone on (b)(6) 2021, the nurse stated that they switched the arctic sun device with another one and was able to complete therapy with no further issues.
The nurse stated that the first arctic sun was labeled for biomed.
|