• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a rotator cuff surgery the device broke inside the patient.The broken piece wasn`t retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-1927bcf-3-1 inserter was received for investigation.Visual inspection identified heavy suture breakage at the distal tip of the inserter.No damage was identified to the inserter itself.The method used to prepare the bone, as well as the bone quality encountered, were not provided.As the implant remains in the patient, no assessment could be made as to the condition of the implant.The cause remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, BIOCOMPOSITE CORKSCREW
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12737955
Report Number1220246-2021-03870
Device Sequence Number11959134
Product Code MAI
UDI-Device Identifier00888867096790
UDI-Public00888867096790
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K082810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberSUTURE ANCHOR, BIOCOMPOSITE CORKSCREW
Device Catalogue NumberAR-1927BCF-3-1
Device Lot Number13015328
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2021
Supplement Date Received by Manufacturer10/14/2021
Initial Report FDA Received Date11/02/2021
Supplement Report FDA Received Date11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-