| Model Number |
IPN042498 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the cuff of proseal #4 did not inflate when a nurse tried to inflate it after intubation".No patient injury or harm reported.Patient condition reported as "fine".It was reported that device was replaced with a new one.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that the airway tube was yellowish due to multiple uses.The device was functionally tested and it was able to be inflated and deflated with a syringe.No blockage was found on the red plug, and air inside the device could be released via the red plug.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.There were no issued found with the returned device.
|
| |
|
Event or Problem Description
|
|
It was reported that"the cuff of proseal #4 did not inflate when a nurse tried to inflate it after intubation".No patient injury or harm reported.Patient condition reported as "fine".It was reported that device was replaced with a new one.
|
| |
|