Model Number NSLG2C25
Medical Device Problem Codes Difficult to Open or Close (2921); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: v9527x.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2c25 device was returned with the bottom portion of the i blade out of the lower jaw channel; the lower jaw channel was damaged.In addition, no damaged to the jaw was observed.The device was connected to the generator and it was recognized.Because of the condition of the device not all functional testing could be performed with the generator.The jaw was unable to cycle open and close.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an exploratory laparotomy procedure that the device ¿jaws would not open up, it broke¿.Unknown as to how the procedure was completed.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 CURVED JAW 25CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key12738212
Report Number3005075853-2021-06604
Device Sequence Number4238357
Product Code GEI
UDI-Device Identifier10705036014133
UDI-Public10705036014133
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Model NumberNSLG2C25
Device Catalogue NumberNSLG2C25
Device Lot NumberV9542Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2021
Initial Report FDA Received Date11/02/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR.
-
-