Model Number STRIP, CVS 100CT KETONE
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Sections with additional information as of 01-dec-2021: device evaluated by manufacturer.Updated fda¿s type, findings and conclusions codes.Ketone strips were returned for evaluation.Defect found on returned strips: physical defect (discolored grey pads).Root cause: rc-061: storage outside specifications.
 
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Ketone test strips were returned for evaluation.Investigation in process note: manufacturer made several attempts to contact customer to ensure the replacement products resolved the initial concern - unable to establish contact with customer.
 
Event or Problem Description
Consumer reported complaint for physical defect of the ketone test strips.Customer stated the ketone test strip pads are grey/brown in color.The package had not been open or damaged when received by the customer.The product is not stored according to specification (bathroom).The test strip lot manufacturer¿s expiration date is 06/30/2022 and test strips were opened one month ago.The customer feels well and did not report any symptoms.No medical attention associated with the use of the product was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KETONE
Common Device Name
NITROPRUSSIDE, KETONES (URINARY, NON-QUANT.)
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key12738523
Report Number1000113657-2021-00693
Device Sequence Number20037662
Product Code JIN
UDI-Device Identifier00021292008178
UDI-Public021292008178
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2022
Device Model NumberSTRIP, CVS 100CT KETONE
Device Lot NumberAX601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2021
Is the Reporter a Health Professional? No
Distributor Facility Aware Date10/07/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer11/09/2021
Initial Report FDA Received Date11/02/2021
Supplement Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-