| Model Number |
STRIP, CVS 100CT KETONE |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 01-dec-2021: device evaluated by manufacturer.Updated fda¿s type, findings and conclusions codes.Ketone strips were returned for evaluation.Defect found on returned strips: physical defect (discolored grey pads).Root cause: rc-061: storage outside specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Ketone test strips were returned for evaluation.Investigation in process note: manufacturer made several attempts to contact customer to ensure the replacement products resolved the initial concern - unable to establish contact with customer.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for physical defect of the ketone test strips.Customer stated the ketone test strip pads are grey/brown in color.The package had not been open or damaged when received by the customer.The product is not stored according to specification (bathroom).The test strip lot manufacturer¿s expiration date is 06/30/2022 and test strips were opened one month ago.The customer feels well and did not report any symptoms.No medical attention associated with the use of the product was reported.
|
| |
|
Search Alerts/Recalls
|