| Model Number |
50-501XL |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that during a myosure procedure on an unknown date , metal shavings were found inside the uterine cavity , no other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Myosure disposable received and tested.Visual inspection was performed and had no kinks or missing components.The outter cannula and the blade were corroded.The mechanical testing was executed and could reciprocate without obstruction.The dissection testing shows the blade was broke and the tip of the cannula shows pieces missing.Hence the complaint can be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|
| |
|
Search Alerts/Recalls
|