ETHICON INC. ULTRAPRO ADVANCED; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number UPA31015
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Evaluation: one unopened sample of product was returned for evaluation.Upon visual inspection of the returned sample, the foil was inspected and multiples holes in the cavity and were noted.In addition, the foil was observed with marks that appears to be caused by a pen.Due to the damages found on the device, a possible cause for these conditions is due to improper handling during transit or storage.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional information was requested and the following was obtained: please explain in detail how was the primary package defective? was there any hole, tear, or puncture that compromises sterility? any open or incomplete seal? foil packaging of mesh shows a series of punctiform indentations.It is unclear, whether the foil actually perforated.Please clarify if there were any adverse patient consequences due to this event? if yes, please explain in detail: issue was detected pre op, therefore no harm to patient how was procedure successfully completed? with new mesh.Please provide status of product return: product is in transit to loc.
 
Event or Problem Description
It was reported that a patient underwent a hernia repair on an unknown date in 2021 and the mesh was used.Prior to surgery, it was found that the primary packaging was defective.It was reported that the foil packaging of the mesh showed a series of punctiform indentations, but it was unclear whether the foil actually perforated.A like device was used to successfully complete the procedure.There were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 05/17/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was provided for h3 evaluation: one unopened sample of product was returned for evaluation.Upon visual inspection of the returned sample, the foil was inspected and multiples holes were noted and the sterile barrier was compromised.The packaging was opened to confirm that the holes were noted to be from the outside in causing sterility breach.In addition, the foil was observed with marks that appear to be caused by a pen.Due to the damages found on the device, the complaint is confirmed, that a possible cause for these conditions is due to improper handling during transit or storage.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/11/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAPRO ADVANCED
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-GMBH
robert-koch strasse 1
p.o. box 1409
norderstedt D2284 1
GM   D22841
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key12740915
Report Number2210968-2021-10660
Device Sequence Number19858539
Product Code FTL
UDI-Device Identifier10705031236998
UDI-Public10705031236998
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K070224
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberUPA31015
Device Catalogue NumberUPA31015
Device Lot NumberQLBCDKD0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2021
Supplement Date Received by Manufacturer04/18/2022
11/09/2022
Initial Report FDA Received Date11/02/2021
Supplement Report FDA Received Date05/18/2022
11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-