| Model Number |
IPN915491 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "during a procedure of checking devices, the pediatric nurse checked a laryngoscope handle to make sure it was properly working.She snapped the handle during few seconds on a blade.The light produced was not very bright.She removed the blade and decided to change the batteries of the handle.When she handled the bottom of the laryngoscope handle to unscrew it and remove the batteries, she noticed that this area was burning hot and quickly unhanded the handle.After few minutes it was still hot but no longer burning hot".No patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The actual device was returned and sent to the manufacturing site for investigation.The manufacturing site reports "there were no issues identified, with the actual complaint sample.And the complaint device is meeting manufacturing and design /drawing release specification.And found acceptable in functional testing.And the complaint device is performing as intended for use.Therefore, it is concluded that this complaint is not justified".
|
| |
|
Event or Problem Description
|
|
It was reported, "during a procedure of checking devices.The pediatric nurse checked a laryngoscope handle to make sure it was properly working.She snapped the handle, during few seconds on a blade.The light produced was not very bright.She removed the blade and decided to change the batteries of the handle.When she handled the bottom of the laryngoscope handle to unscrew it and remove the batteries, she noticed that this area was burning hot and quickly unhanded the handle.After few minutes, it was still hot, but no longer burning hot".No patient involvement reported.
|
| |
|
Search Alerts/Recalls
|