| Model Number |
1588020122 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a blurry image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: sticky cord/ image ranged from obstructive to blurry.The failure(s) identified in the investigation is consistent with the complaint record.Root cause: film and rainbow on the windows.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Search Alerts/Recalls
|