| Model Number |
OJR412 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint ojr412b is currently en route to fisher & paykel healthcare (f&p) in (b)(4) for evaluation.We will provide a follow up report upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr412b optiflow junior 2 nasal cannula had detached from the swivel grip tube joint.There was no reported patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint ojr412 optiflow junior 2 nasal cannula was returned to fisher & paykel healthcare (f&p) where it was visually inspected.Results: visual inspection of the returned device revealed that one side of the tube of the ojr412 optiflow junior 2 nasal cannula was found detached from the swivel grip.Conclusion: we are unable to determine the cause of the reported event.However, the observed damage was most likely caused by the tubing being pulled.All optiflow junior 2 cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 cannula.They also state the following: appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death.Do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).Ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in france reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr412b optiflow junior 2 nasal cannula had detached from the swivel grip tube joint.There was no reported patient consequences.
|
| |
|