| Model Number |
15 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's main keypad to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device's power button was functioning intermittently.In this state the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported with the event.
|
| |
|