MEDTRONIC MEXICO 13 GA TAPERED CANNULA, TROCAR STYLET, BEVEL STYLET; DEVICE, PERCUTANEOUS, BIOPSY Back to Search Results
Model Number VPB13D
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via manufacturer representative regarding a patient for unknown spinal therapy.It was reported that when removing cannula from patient (after retrieving biopsy) the top of the cannula came detached.There was patient involved in the event and no further complications or symptoms were reported.Additional information was received via manufacturer representative that there were no fragments left inside the patient.Hcp was able to remove the detached cannula using a surgical clamp.The product came in contact with the patient.Pre-op diagnosis and procedure involved in the event was bone biopsy.
 
Additional Manufacturer Narrative
H3: product analysis: part# vpb13, lot# 221828512 visual optical visual and optical inspection confirmed the cannula was returned with the hub pulled from the shaft.On the cannula, the knurling was present as well as the flair on the backside of the shaft.It appears the shaft was separated from the plastic hub from excessive force.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
13 GA TAPERED CANNULA, TROCAR STYLET, BEVEL STYLET
Common Device Name
DEVICE, PERCUTANEOUS, BIOPSY
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12744710
Report Number9612164-2021-04192
Device Sequence Number11879823
Product Code MJG
UDI-Device Identifier00763000026769
UDI-Public00763000026769
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/05/2023
Device Model NumberVPB13D
Device Catalogue NumberVPB13D
Device Lot Number221828512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2021
Supplement Date Received by Manufacturer02/02/2022
Initial Report FDA Received Date11/03/2021
Supplement Report FDA Received Date02/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-