| Model Number |
45031 |
| Medical Device Problem Codes |
Suction Problem (2170); Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device returned contained an angiojet solent omni.The assembly, supply line, shaft, tip, and spike line were all visually examined for potential damage.The catheter had multiple kinks along the catheter shaft.A functional test was performed.The pump was placed into the ultra drive console and the device primed with no complication.The device was run for a total time of 90 seconds in thrombectomy mode.The device stayed within the expected range of the product, with a pressure of 10.646 kpsi.There is no confirmation for the complaints allegation of shaft perforation or loss of aspiration.
|
| |
|
Event or Problem Description
|
|
It was reported that loss of aspiration occurred.An angiojet solent omni catheter was used for the thrombectomy procedure.During the procedure, it was reported there was a hole in the shaft of the catheter causing diminished suction power.The device continued to pump saline when aspiration was lost and the procedure was completed successfully.No patient complications were reported.
|
| |
|