• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. 18IN. (46CM) DISPOSABLE CUFF W/PLC - DP, SB; TOURNIQUET, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. 18IN. (46CM) DISPOSABLE CUFF W/PLC - DP, SB; TOURNIQUET, PNEUMATIC Back to Search Results
Model Number N/A
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event or Problem Description
It was reported that the tourniquet cuff was leaking air.There was no spontaneous deflation.The cuff was unable to be used so another one was opened for use.The event occurred before surgery.No adverse event was reported as a result of this malfunction.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).When connected to a known working ats in the lab, the ats had to keep pumping air to keep the cuffs inflated, confirming a leak.A bubble test determined the leak was located where the pvc tubing connects to the right angle connectors of all 4 of the cuffs.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the tourniquet cuff was leaking air.There was no spontaneous deflation.The cuff was unable to be used so another one was opened for use.The event occurred before surgery.This device was brand new and just removed from its sterile packaging.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
18IN. (46CM) DISPOSABLE CUFF W/PLC - DP, SB
Common Device Name
TOURNIQUET, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12745801
Report Number0001526350-2021-01221
Device Sequence Number6078466
Product Code KCY
UDI-Device Identifier00889024380578
UDI-Public(01)00889024380578(17)240527(10)34237275
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2024
Device Model NumberN/A
Device Catalogue Number60707010300
Device Lot Number34237275
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/20/2021
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date11/03/2021
Supplement Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-