MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES; BLOOD CULTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES; BLOOD CULTURING SYSTEM Back to Search Results
Model Number 245122
Device Problem Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that 7 bd bbl¿ mgit¿ mycobacteria growth indicator tubes had no label on them.The following information was provided by the initial reporter, translated from chinese to english: "mgit tube has no identifiable label.".
 
Manufacturer Narrative
H.6.Investigation: material 245122 is manufactured by rehydrating the media components with usp purified water, and thoroughly mixing until a homogeneous solution is obtained.The tubes are filled, capped, torqued, and then labeled by machine per standard operating procedure (sop).The tubes are terminally autoclaved in an air over pressure (aop) autoclave, per manufacturing instructions, using a validated cycle.Post autoclaving, tubes are packaged into final shipping configurations.The batch history record review for batch 1063895 was satisfactory per internal procedures.Filling, labeling and packaging processes were within specifications.Qc inspection and testing were satisfactory at time of release.The complaint history was reviewed, and three other complaints have been taken on this batch for labeling.Retention samples from batch 1063895 (100 tubes) were available for inspection.No labeling defects were observed in 100/100 retention samples.All retention tubes had one tube label that had legible batch information and a scannable barcode.No photos were received to assist with the investigation.No returns were received to assist with the investigation.This complaint cannot be confirmed.Bd will continue to trend complaints for labeling.
 
Event Description
It was reported that 7 bd bbl¿ mgit¿ mycobacteria growth indicator tubes had no label on them.The following information was provided by the initial reporter, translated from chinese to english: "mgit tube has no identifiable label".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES
Type of Device
BLOOD CULTURING SYSTEM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12745943
MDR Text Key279883454
Report Number1119779-2021-01751
Device Sequence Number1
Product Code MDB
UDI-Device Identifier30382902451229
UDI-Public30382902451229
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K974883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/02/2022
Device Model Number245122
Device Catalogue Number245122
Device Lot Number1063895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-