Model Number SGC0701
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation (2001)
Event Date 10/15/2021
Event Type  Injury  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report the atrial septal defect requiring intervention. It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4+. An intra-aortic balloon pump was placed prior to the start of the procedure. The clip delivery system (cds) was advanced to the mitral valve however the leaflets could not be grasped therefore the cds was removed. A second cds was advanced and deployed, reducing mr to 2+. Once the steerable guide catheter (sgc) was removed a right to left shunt was observed therefore an amplatzer asd device was successfully placed. The balloon pump was then exchanged for an impella device for further post-op support, and it was then observed that the patient blood pressure and vitals dropped. Attempts were made to improve the patients condition however the patient expired. Per the physician, the patient death was not related to the mitraclip devices. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG4 STEERABLE GUIDING CATHETER
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12746582
MDR Text Key279919539
Report Number2024168-2021-09935
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number10726R148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/03/2021 Patient Sequence Number: 1
Treatment
1 IMPLANTED MITRACLIP; INTRA-AORTIC BALLOON PUMP
-
-