| Model Number |
TSM-275 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: visual analysis of the returned device identified the device was broken shell and missing shell.A weight test of the device was verified and the device underweight.A microscopic analysis was performed which identified broken sharp.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp edge broken in the side posterior assessed as unidentified (tear) opening/.Missing shell assessed as inconclusive.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side rupture.The device has been explanted.
|
| |
|