| Model Number |
APB-4-10-3D-SS |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report an axium detachment failure.No patient injury was identified.The unruptured aneurysm was on the right ic-pc5mm with an amorphous shape.The max diameter of the aneurysm was 5 mm with a neck diameter of 4 mm.The blood flow speed and tortuosity was normal.The procedure was started, according to the plan that a stent would be deployed after the coil embolization.Although the reported device was used after 360 soft of 5×15 was placed on a frame, it could not be detached; it was detached where about 4 mm of the tail part was pulled out from the aneurysm.Since the stent deployment was planned finally, the embolization was continued.The procedure was finished by holding down the tail part of the coil that deviated in the parent blood vessel with the stent as planned.Several attempts were made to detach the coil.The doctor didn¿t attempt.To detach the coil using another instant detacher, a manual attempt was made.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating there was nothing in particular that led to the coil non-detachment.There seemed to be no problem in particular by visual inspection.
|
| |
|
Search Alerts/Recalls
|