• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752201
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A product sample has been received by the manufacturer and it is awaiting evaluation.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported the aspiration tube occluded with two different cassette pak during a cataract procedure.The procedure was completed with the third replacement.There was no harm to the patient.This is two of two reports for this event.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.The wet fluidics management system (fms), in the tray was visually inspected and functionally tested.Both irrigation and aspiration luers were intact.Flare shape was observed, in the blue socket fitting at the aspiration tubing insertion end.The sample primed and tuned with a handpiece successfully.The sample could be recognized and the service data could be retrieved from the console.The sample passed functionality testing.The root cause of the customer's complaint could not be established, since the returned sample met specifications.No contributing factors could be identified, that could cause the reported complaint.After the investigation of this complaint.It has been determined, that this sample met specifications.Therefore, no action will be taken at this time.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences, as is deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12746802
MDR Text Key279944291
Report Number1644019-2021-00684
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657522019
UDI-Public00380657522019
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue Number8065752201
Device Lot Number2463336H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-