| Model Number |
777726 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced stent dislodgement, migration, pain, discomfort and infection by ureteral stent.No medical intervention was reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced stent dislodgement, migration, pain, discomfort and infection by ureteral stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.No sample was returned for evaluation.It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.A potential root cause for this failure could be "material selection" or "part geometry".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "potential complications associated with retrograde/antegrade positioning of indwelling ureteral stents include the following: edema.Stone formation.Peritonitis.Extravasation.Ureteral reflux.Stent dislodgement, fragmentation, migration, occlusion.Fistula formation.Loss of renal function.Hemorrhage.Pain/discomfort.Stent encrustation.Hydronephrosis.Perforation of kidney, renal pelvis, ureter and/or bladder.Ureteral erosion.Infection.Urinary symptoms." "with any ureteral stent, migration is a possible complication which could require medical intervention for removal.Selection of too short a stent may result in migration." "determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further review, bd has determined that this mdr was reported in error as it was found to be a duplicate of an event previously reported.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced stent dislodgement, migration, pain, discomfort and infection by ureteral stent.
|
| |
|
Search Alerts/Recalls
|