Manufacturer's investigation conclusion: thrombus was confirmed through the evaluation of heartmate ii lvas, serial number (b)(6).A direct correlation to the heartmate ii lvas, serial number (b)(6), and the reported events could not conclusively be determined through this evaluation.The submitted log file contained data from 20:02:13 on (b)(6) 2021 through 13:57:48 on (b)(6) 2021, per the timestamps.Throughout the captured data, elevations in pump power as high as 18.2 w were captured.Corresponding elevations in pump flow, as high as 12.0 lpm, were also captured.The pump operated above the low speed limit for the duration of the log file.A specific cause for these findings could not be determined through this evaluation.(b)(6) was returned assembled with the driveline (dl) cut approximately 2.5¿¿ from the pump housing and the distal end of the driveline was not returned.The sealed inflow conduit and sealed outflow graft were not returned.The proximal and distal-sides of the pump stators showed no evidence of deposition or thrombus.The outlet elbow was returned attached to the outlet port.Upon disassembly of (b)(6) , visual inspection of the rotor bearing ball revealed a small, denatured thrombus.The thrombus showed evidence of laminated layering indicating that it formed in this location over an undetermined amount of time.The deposition found in the pump could have potentially contributed to the reported hemolysis.The deposition could have also created additional resistance on the spinning rotor, potentially contributing to the reported elevations in pump power.Upon removal of the observed deposition, the device was cleaned.The pump¿s bearings, rotor and blood-contacting surfaces were examined under a microscope and no anomalies were observed.Electrical continuity testing of the driveline revealed no discontinuities or shorts.The pump was reassembled and functionally tested under normal operating conditions using a mock circulatory loop.The data retrieved from the testing revealed normal pump power consumption and pressure values, comparable to what was recorded during the manufacturing process and the device functioned as intended.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The implant kit was shipped on 27dec2017.The heartmate ii lvas ifu and the heartmate ii patient handbook are currently available.Section 1 of the ifu, ¿introduction¿, lists hemolysis and device thrombosis as adverse events that may be associated with the use of the heartmate 3 left ventricular assist system.Section 6 of the ifu, ¿patient care and management¿, also lists thromboembolism as a potential late postimplant complication.Section 1 and section 6 of the ifu also outline indications of pump thrombosis and how to respond to such events.Section 6, under ¿anticoagulation¿, provides information regarding the recommended anticoagulation therapy and inr range, as well as suggested anticoagulation modifications.Section 1 also addresses all pump parameters including pump power.This section explains that pump power is a direct measurement of motor voltage and current; therefore, changes in pump speed, flow, or physiological demand can affect pump power.This section also notes that any increase in power not related to increased flow causes erroneously high flow readings.Additionally, section 6, under ¿pump performance monitoring¿, explains that pump performance is sensitive to changes in systemic vascular resistance and left ventricular filling.No further information was provided.The manufacturer is closing the file on this event.
|