| Model Number |
SCM-275 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The event of capsular contracture, baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation is capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture, baker grade iii" against left device.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: weight to the spec and creases fold.No other observations observed in the device.Based on the device analysis the final assessment is: - no issues found related with the manufacturing process.
|
| |
|