OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-3
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2021
Event Type  malfunction  
Event Description
The olympus field service engineer became aware that "when the high flow insufflation unit switched on, it immediately and continuously restarts." the event was found after an unspecified endo-procedure during bedside cleaning.No patient injury or harm was reported.
 
Manufacturer Narrative
The device was returned for inspection and the reported the unit auto-restarts continuously due to main board malfunction.Additionally, the flow rate readings are not within the standard due to electric flow unit is faulty and the chassis is corroded.The investigation is ongoing; therefore, the root cause of the reported issue/malfunction cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide a correction for g2 and additional information based on the legal manufacturer's final investigation.G2 - checked foreign and 'other' to add the country singapore.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 14 years since the subject device was manufactured.Based on the results of the investigation, the cause of the device continuously restarting was from a defective main board.The flow rate readings not being within standard was caused by a defective electropneumatic proportional valve.The specific root cause of the defective main board and electropneumatic proportional valve could not be determined at this time.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12749135
MDR Text Key283610512
Report Number8010047-2021-14031
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170140280
UDI-Public04953170140280
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K014166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-