Model Number UHI-3 |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2021 |
Event Type
malfunction
|
Event Description
|
The olympus field service engineer became aware that "when the high flow insufflation unit switched on, it immediately and continuously restarts." the event was found after an unspecified endo-procedure during bedside cleaning.No patient injury or harm was reported.
|
|
Manufacturer Narrative
|
The device was returned for inspection and the reported the unit auto-restarts continuously due to main board malfunction.Additionally, the flow rate readings are not within the standard due to electric flow unit is faulty and the chassis is corroded.The investigation is ongoing; therefore, the root cause of the reported issue/malfunction cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide a correction for g2 and additional information based on the legal manufacturer's final investigation.G2 - checked foreign and 'other' to add the country singapore.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 14 years since the subject device was manufactured.Based on the results of the investigation, the cause of the device continuously restarting was from a defective main board.The flow rate readings not being within standard was caused by a defective electropneumatic proportional valve.The specific root cause of the defective main board and electropneumatic proportional valve could not be determined at this time.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|