ETHICON ENDO-SURGERY, LLC. BASX BLADELESS TROCAR 5MM; LAPAROSCOPE, GENERAL, & PLASTIC SURGERY
|
Back to Search Results |
|
| Model Number |
TB5LT |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.An analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.As part of our quality process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, before used on the patient, noted the device was broken (sleeve issue).Another device was used to complete the surgery.There was no patient consequence reported.No additional information could be provided.
|
| |
|
Search Alerts/Recalls
|
|
|