| Model Number |
7211004 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during use instruments outside patient, there was a fluid leaked from the dyonics 25 cassette.The cassette was missing the transducer membrane and the main body of the pump broke down due to the fluid leakage.The procedure was completed with a s+n back-up device.A significant delay was reported.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.A visual inspection revealed that the cassette is missing two transducer membranes.A functional evaluation was not performed because the cassette would leak fluids due to the missing transducer membranes.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined.Factors that could have contributed to the failure include pressurizing a cassette prior to insertion into the controller.No containment or corrective actions are recommended at this time.
|
| |
|