• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-3
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 10/10/2021
Event Type  Death  
Manufacturer Narrative
There was no device malfunction.The device was received for evaluation and successfully passed testing, confirming proper functionality of the device and sensors.Available log files were retrieved and analyzed which showed the device was performing as designed and intended.A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.The nxstage system one user guide outlines risks associated with performing hemodialysis therapy and warns that treatment should only be performed by a trained and qualified person who must respond promptly to harmful conditions during treatment.Udi:(b)(6).
 
Event Description
A report was received on 12 oct 2021 from the home therapy nurse (htn) of a (b)(6) male with a medical history including diabetes and end stage renal disease, who stated the patient expired at an unspecified time during a home hemodialysis treatment on (b)(6) 2021.Additional information was received on 13 oct 2021 from the htn who stated the patient treated for approximately 3 hours before being found by a family member and 911 was called.Although requested, a cause of death was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
MDR Report Key12750797
MDR Text Key280086425
Report Number3003464075-2021-00065
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-3
Device Catalogue NumberCHRONIC HI-FLOW CYCLER
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/12/2021
Initial Date FDA Received11/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age61 YR
Patient Weight122
-
-