• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEC,BLDE,LNG,EXT,INSULATD,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 138110
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event of ¿insufficient heatseal¿ is confirmed.Two 138110 returned unopened in original packaging.The lot number of the devices, 202004271 was verified.A visual inspection found an open hole in the packaging of one device caused by the electrode encroaching into the seal.The second device was also encroaching into the seal of the packaging.The second device without an obvious breach was dye leak tested, which indicated an insufficient heat seal, the dye leaked.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A 2 year lot history review was conducted and found this is the only event with a quantity of 2 units for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 77 complaints, regarding 378 devices, for this device family and failure mode.During this same time frame 1,972,470 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: sterility is guaranteed unless package has been opened, broken or damaged.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
During incoming inspection, the distributor rejected this device, 138110, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEC,BLDE,LNG,EXT,INSULATD,40
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
MDR Report Key12751420
Report Number1320894-2021-00400
Device Sequence Number4558020
Product Code GEI
UDI-Device Identifier30653405003452
UDI-Public(01)30653405003452(17)250426(10)202004271
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K943542
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/26/2025
Device Model Number138110
Device Catalogue Number138110
Device Lot Number202004271
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2021
Initial Report FDA Received Date11/04/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device N
Patient Sequence Number1
-
-