| Model Number |
406137 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During atrial fibrillation procedure, an air leak occurred.After placing the sheath, prior to inserting catheters, air was aspirated into a syringe that connected to the extension port of the sheath.Several aspirations were attempted but the air was still aspirated.The sheath set was replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f fast-cath introducer sheath was received for evaluation. the sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
|
| |
|