| Model Number |
2760 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter city: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the gauge needle did not moved down.An encore 26 advantage kit was selected for use.During the procedure, it was noted that the meter part of the indeflator was unusual as the gauge needle did not move down when deflation was performed after pressure was applied to 24 atmospheres.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the gauge needle did not move down.An encore 26 advantage kit was selected for use.During the procedure, it was noted that the meter part of the indeflator was unusual as the gauge needle did not move down when deflation was performed after pressure was applied to 24 atmospheres.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter city: (b)(6).Device evaluated by manufacturer: the device was returned for analysis.The gauge needle was at 26 atmospheres (atm) when received.A functional test of the complaint device was carried out using a bsc stopcock.The unit failed all gauge accuracy tests due to inaccurate readings at 13 atm and 26 atm.
|
| |
|