| Model Number |
SCX-340 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Lab analysis summary: analysis of the returned device identified: weight to the spec and opening on anterior.A visual and microscopic analysis was performed which identified striated opening on anterior and crease flat.Base on the device analysis the final assessment is: striated opening assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Search Alerts/Recalls
|