• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C8750
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2021
Event Type  malfunction  
Manufacturer Narrative
The patient was an adult.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified volume of blood was leaking at the insertion site while using a clearlink system y-type blood/solution set.This was observed during blood transfusion to a patient.Upon inspection, the tubing separated off.Another blood transfusion bag was readily available for the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h6 and h10.H10:the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which did not identify any abnormalities that could have contributed to the reported condition.The reported condition is not clearly observed via the provided photograph and due to the nature of the returned sample no additional testing could be performed.Therefore, the cause of the reported condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12753450
MDR Text Key280201467
Report Number1416980-2021-06610
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8750
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BLOOD TRANSFUSION BAG.; BLOOD TRANSFUSION BAG
-
-