• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ATELLICA IM BR CA 27.29; BR IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ATELLICA IM BR CA 27.29; BR IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
Customer observed discordant elevated atellica im 1600 br (br 27.29) results for two patient samples compared to repeat testing and historical results.The results were reported to the physician and were not questioned.There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant elevated results.
 
Additional Manufacturer Narrative
A us customer reported elevated atellica im 1600 br (br 27.29) results for two patient samples compared to repeat testing and historical results.Samples are run from aliquot tubes and retrieved from high-volume storage for repeat testing.Siemens is investigating.The instructions for use (ifu) states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." the limitations section of the instructions for use states the following: "note do not interpret levels of ca 27.29 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed breast carcinoma frequently have levels of ca 27.29 within the range observed in healthy individuals.Additionally, elevated levels of ca 27.29 can be observed in patients with nonmalignant diseases.Measurements of ca 27.29 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.".
 
Additional Manufacturer Narrative
The initial mdr 1219913-2021-00475 was filed on november 4, 2021.A us customer reported additional elevated atellica im atellica im 1600 br (27.29) results for two patient samples compared to repeat testing.Atellica im br ca 27.29.Units of measure: u/ml.Sid: (b)(6), result: 30.86, date: (b)(6) 2021; repeated, 46.36, (b)(6) 2021; (b)(6), 33.92, (b)(6) 2021; repeated, 35.83, (b)(6) 2021; repeated, 40.03, (b)(6) 2021; repeated, 54.18, (b)(6) 2021; repeated, 31.49, (b)(6) 2021.Siemens is investigating.
 
Additional Manufacturer Narrative
The initial mdr 1219913-2021-00475 was filed on november 4, 2021 and supplemental 1 mdr was filed on january 7, 2022.A us customer observed elevated, non reproducible atellica im 1600 br (27.29) lot: 257 results from patient samples.Samples were repeated on different atellica im analyzers and at times the testing lasted for more than two days post initial result.The atellica im br instructions for use (ifu) (11206410) indicates not to use samples that have been stored at room temperature for longer than 8 hours and to freeze samples at -20°c if the assay is not completed within 48 hours.The customer is aware of these limitations but they have validated the storage of br samples for longer duration.A review of data did not show a consistent cause for the difference in patient sample recovery between runs.In some instances the difference could be attributed to the br calibration, but not in every case.In some instances the qc recovery from a br primary readypack trended with the patient recovery off of the same pack indicating an issue with the pack, but not in every case.Instances of improper reagent handling and storage were noted at the customer's warehouse, during the time kit lot: 257 was stored there, and at the customer site, though not specifically with br packs.However, there is no direct evidence that the handling and storage of br reagents contributed to the issue.During the service at the customer site, solid phase particles were observed at the top of the ready packs which has the potential to impact sample recovery.Atellica im br kit lot: 257, expired 2021/11/10, so no internal studies were done with the lot, and the customer moved to atellica im br kit lot: 259.The cause(s) of the variability of patient sample recovery run to run seen by the customer when using atellica im br kit lot: 257 could not be determined but siemens cannot rule out pre-analytical factors, a sample issue, normal assay performance, or reagent storage/handling.Based on the investigation, no product problem was identified.The customer is operational.No further action is required.In section h6, investigation finding and investigation conclusion codes were updated based on the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA IM BR CA 27.29
Common Device Name
BR IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS, INC
333 coney street
east walpole MA 02032
Manufacturer Contact
louise mclaughlin
333 coney street
east walpole, MA 02032
7818564812
MDR Report Key12754995
Report Number1219913-2021-00475
Device Sequence Number13800081
Product Code MOI
UDI-Device Identifier00630414598055
UDI-Public00630414598055
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K982680
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2021
Device Model NumberN/A
Device Catalogue Number10995478
Device Lot Number257
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2021
Supplement Date Received by Manufacturer12/13/2021
03/03/2022
Initial Report FDA Received Date11/04/2021
Supplement Report FDA Received Date01/07/2022
03/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/10/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-