Model Number 383594
Medical Device Problem Code Material Perforation (2205)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd nexiva¿ diffusics¿ closed iv catheter system, the device experienced the needle perforating through the catheter.This event occurred four times.The following information was provided by the initial reporter.The customer stated: the catheter has is damage by the point of the needle where it is introduce to the body customer said it had happened 4 times and he has one of the product to return for investigation.
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd nexiva¿ diffusics¿ closed iv catheter system, the device experienced the needle perforating through the catheter.This event occurred four times.The following information was provided by the initial reporter.The customer stated: the catheter has is damage by the point of the needle where it is introduce to the body customer said it had happened 4 times and he has one of the product to return for investigation.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2021-10-18 investigation summary: bd received 1 sample submitted for evaluation.The reported issue was not confirmed upon inspection of the sample, since no defect was observed when the tip of the needle was observed with a microscope.Since the reported failure was not confirmed a root cause related to our manufacturing process cannot be established.A device history record review showed no non-conformances associated with this issue during the production of this batch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA¿ DIFFUSICS¿ CLOSED IV CATHETER SYSTEM
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12755289
Report Number9610847-2021-00534
Device Sequence Number4614290
Product Code FOZ
UDI-Device Identifier30382903835943
UDI-Public30382903835943
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K123734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2024
Device Model Number383594
Device Catalogue Number383594
Device Lot Number1015996
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2021
Supplement Date Received by Manufacturer11/08/2021
Initial Report FDA Received Date11/04/2021
Supplement Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-