| Model Number |
3544650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a (b)(6) caucasian female patient underwent a revision breast augmentation with two 650cc mentor memorygel breast implant prostheses and experienced bilateral rupture postoperatively.The diagnosis of the right-sided rupture was confirmed based on mammogram.A healthcare professional reported that the left-sided device is going to be removed due to suspected rupture (due to the age of the implant).As a result, the patient is scheduled to undergo removal and replacement with unspecified gel breast implant on (b)(6) 2022.This report is for the left-sided prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: suspected rupture.Date of explantation: (b)(6) 2022.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 7-mar-2022, mentor received additional information regarding the mammogram.Mammogram was performed on (b)(6) 2021.On 18-mar-2022, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 13-apr-2022, the investigation on the suspected medical device was completed.Investigation summary: mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed parallel lines of shell wear extended from the anterior to the posterior view.Leak testing was performed, in accordance with mentor procedures, and a tear was observed within the shell wear on the posterior view, measuring approximately 0.2 cm.The evaluation determined that the cause of the rupture is consistent with normal wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.As part of the mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
|
| |
|