| Model Number |
25002112E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
10/13/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in process.When the investigation is complete, a supplemental mdr will be submitted.Multiple mdrs have been submitted for this adverse event: #3013450937-2021-00254.#3013450937-2021-00256.#3013450937-2021-00257.#3013450937-2021-00258.#3013450937-2021-00259.#3013450937-2021-00260.#3013450937-2021-00261.#3013450937-2021-00262.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient developed an alleged infection and underwent a revision surgery on (b)(6) 2021.The surgeon performed a washout and poly swap.During the revision surgery, the following eleos implants were revised: resurfacing femur axial pin, poly spacer, and tibial hinge component.The following eleos implants remain implanted from the original surgery: cemented stem extension, biogrip collar, collar locking ring, resurfacing femur, cemented stem, and proximal tibia.No additional information regarding this adverse event has been provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient developed an alleged infection and underwent a revision surgery on (b)(6) 2021.The surgeon performed a washout and poly swap.During the revision surgery, the following eleos implants were revised: resurfacing femur axial pin, poly spacer, and tibial hinge component.The following eleos implants remain implanted from the original surgery: cemented stem extension, biogrip collar, collar locking ring, resurfacing femur, cemented stem, and proximal tibia.No additional information regarding this adverse event has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.
|
| |
|
Search Alerts/Recalls
|