• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; MODULAR COLLAR LOCKING RING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; MODULAR COLLAR LOCKING RING Back to Search Results
Model Number HR-30000-03M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in process.When the investigation is complete, a supplemental mdr will be submitted.Multiple mdrs have been submitted for this adverse event: #3013450937-2021-00254, #3013450937-2021-00255, #3013450937-2021-00256, #3013450937-2021-00257, #3013450937-2021-00258, #3013450937-2021-00260, #3013450937-2021-00261, #3013450937-2021-00262.
 
Event or Problem Description
It was reported that the patient developed an alleged infection and underwent a revision surgery on (b)(6) 2021.The surgeon performed a wahout and poly swap.During the revision surgery, the following eleos implants were revised: resurfacing femur axial pin, poly spacer, and tibial hinge component.The following eleos implants remain implanted from the original surgery: cemented stem extension, biogrip collar, collar locking ring, resurfacing femur, cemented stem, and proximal tibia.No additional information regarding this adverse event has been provided.
 
Event or Problem Description
It was reported that the patient developed an alleged infection and underwent a revision surgery on (b)(6) 2021.The surgeon performed a washout and poly swap.During the revision surgery, the following eleos implants were revised: resurfacing femur axial pin, poly spacer, and tibial hinge component.The following eleos implants remain implanted from the original surgery: cemented stem extension, biogrip collar, collar locking ring, resurfacing femur, cemented stem, and proximal tibia.No additional information regarding this adverse event has been provided.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
MODULAR COLLAR LOCKING RING
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey rd
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC.
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey rd
parsippany, NJ 07054
MDR Report Key12755827
Report Number3013450937-2021-00259
Device Sequence Number3133956
Product Code KRO
UDI-Device IdentifierB278HR3000003M0
UDI-Public+B278HR3000003M0
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K203815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHR-30000-03M
Device Catalogue NumberHR-30000-03M
Device Lot Number83914R-050
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2021
Supplement Date Received by Manufacturer01/06/2022
Initial Report FDA Received Date11/06/2021
Supplement Report FDA Received Date01/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25001208E, ELEOS TIBIAL POLY SPACER.; P/N 25002112E, ELEOS RESURFACING FEMUR AXIAL PIN.; P/N 2500L002E, ELEOS RESURFACING FEMUR.; P/N HC-09100-03M, ELEOS CEMENTED STEM.; P/N KSC10100E, ELEOS CEMENTED STEM EXTENSION.; P/N PB-2400R-03M, ELEOS COLLAR.; P/N PT-20000-02M, ELEOS PROXIMAL TIBIA.; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age15 YR
Patient SexMale
-
-