MEDACTA INTERNATIONAL SA STEM: MASTERLOC CEMENTLESS TI COATED LAT SIZE 3; CEMENTLESS FEMORAL STEM Back to Search Results
Model Number 01.39.203
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 08-oct-2021: lot 162125: (b)(4) items manufactured and released on 17-jun-2016.Expiration date: 2021-06-06.No anomalies found related to the problem.To date, all items of the same lot have been sold without any similar reported event.Clinical evaluation performed by medacta medical affairs department: 5 years after primary cementless tha the patient reports pain and the radiograph reveals that the femoral stem is broken at neck level.We have one radiograph, taken after the fracture took place, so we cannot draw any conclusion.The event is extremely rare.Analysis of the explants is mandatory in order to understand the mechanisms of the fracture.
 
Event or Problem Description
Revision surgery performed due to stem neck breakage 4 years and 10 months after the primary surgery (stem and head were replaced).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: MASTERLOC CEMENTLESS TI COATED LAT SIZE 3
Common Device Name
CEMENTLESS FEMORAL STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key12758191
Report Number3005180920-2021-00852
Device Sequence Number3310293
Product Code LZO
UDI-Device Identifier07630030885631
UDI-Public07630030885631
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160289
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/06/2021
Device Model Number01.39.203
Device Catalogue Number01.39.203
Device Lot Number162125
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2021
Initial Report FDA Received Date11/07/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/17/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR