| Model Number |
2500CP12E |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned for investigation as it was discarded at the hospital.The investigation is in progress.When the investigation is complete, a supplemental report will be submitted accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to their canal-filling segmental stem loosening.During the revision surgery, the following eleos implants were revised: canal-filling segmental stem, male-female midsection, distal femur, distal femur axial pin, tibial hinge component, and poly spacer.The surgeon explanted the canal-filling segmental stem and placed a cemented segmental stem.No additional information regarding this event has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the segmental stem loosening could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added.G3: date received by manufacturer added.G6: type of report added.H2: follow-up type added.H3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.H10: additional narratives/data.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery on (b)(6) 2021 due to their canal-filling segmental stem loosening.During the revision surgery, the following eleos implants were revised: canal-filling segmental stem, male-female midsection, distal femur, distal femur axial pin, tibial hinge component, and poly spacer.The surgeon explanted the canal-filling segmental stem and placed a cemented segmental stem.No additional information regarding this event has been provided.
|
| |
|