| Model Number |
ESR8202400 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.Coloplast has not been provided any corroborating evidence to verify the information contained in this report.
|
| |
|
Event or Problem Description
|
|
As reported to coloplast, though not verified, tubing fracture by pump.Device was explanted and replaced.No other adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A titan otr pump was received for evaluation.A separation within abrasion was noted on the longer exhaust tube of the pump.This was a site of leakage.Partial separations within abrasion were noted on both exhaust tubes.These were not sites of leakage.Based on examination of the returned product, it was concluded that while in-vivo both exhaust tubes of the pump had overlapped and abraded against one another.This positioning, in combination with device usage over time, may contribute to sufficient stress(s) to separate the longer exhaust tubing.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Event or Problem Description
|
|
According to the available information this inflatable penile prosthesis was implanted on (b)(6) 2014 and revised on (b)(6) 2021 due to a tubing fracture by the pump.
|
| |
|