DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-45 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The distributors service engineer evaluated the iabp unit for the reported batteries that will not charge.To fix the issue service engineer replaced the batteries and the unit now charges, he then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full event site address has been shortened due to field character limits and should read: (b)(6).
|
|
Event Description
|
It was reported that during installation of the cardiosave intra-aortic balloon pump (iabp) there was one of the batteries that was not charging, and not being recognized by the unit.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: g1 (contact person ¿ mfg site).
|
|
Event Description
|
N/a.
|
|