| Model Number |
68LP-175 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The left device has been unilaterally replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified fold creases, one linear opening and yellow particles in device inner surface.A microscopic analysis and leak test were performed which identified one opening crease smooth and wear abrasion.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: -one opening crease smooth in the side posterior assessed as fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device has been explanted and replaced.
|
| |
|