| Model Number |
APB-6-15-3D-SS |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report that the axium prime coil experienced resistance during delivery and "unraveled" in the middle part of the catheter.The coil was damaged/detached early.There was no damage to the delivery pusher, the pusher was not relocated, and no detachment attempts had been made.A continuous flush had been administered.The coil was replaced, and no additional surgical or medical interventions were needed. the patient did not experience any injury.The devices were prepared according to the instructions for use (ifu). the patient's vessel tortuosity was severe, and their blood flow was normal.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating the coil was stretched and retrieved while it was inside the catheter and attached to the pushwire.The resistance occurred in the catheter lumen.The coil was pushed out of the introducer sheath smoothly.The catheter was a non-medtronic product.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|