SMITH & NEPHEW, INC. REDAPT STEM EXTRACTOR UNIV JOINT; PROST, HIP, SEMI-CNSTRND, MET/CER/POLY, CMNTD OR NON-POROUS, UNCMNTD
|
Back to Search Results |
|
| Model Number |
71355107 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal reference number: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the threaded shaft of the redapt stem extractor univ joint is broken.As this was noticed upon field inspection, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.A review of complaint history revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: after further assessment of the information gathered by the manufacturer, it was identified that this event should be re-evaluated for mdr reporting.The original information stated that the threaded shaft of the redapt stem extractor univ joint is broken.As this was noticed upon field inspection, there was no patient involvement.After further clarification and information received, it was determined that the issue does not meet the criteria to be reportable, since the device damage did not lead to a death or serious injury, nor would it be likely to cause or contribute to a death or serious injury if it were to recur.
|
| |
|