Model Number 0998-00-0800-53
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and determined the leak was caused due to user error.Specifically, the fse found a foreign object between the helium tank mating surface and the rubber seal.Once the obstruction was removed the unit passed the helium leak test.The fse performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check, the cardiosave intra-aortic balloon pump (iabp) had a helium leak.Additionally, it was stated that a full tank was installed in the evening and was drained to empty by the following morning.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: d5, g1 (contact person ¿ mfg site).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12760485
MDR Text Key282717756
Report Number2249723-2021-02573
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-