CONCORD MANUFACTURING CRIT-LINE CLIP (CLIC) - USB; ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS Back to Search Results
Model Number CL10041001
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external visual inspection of the returned device showed no signs of physical damage.The verify accuracy test failed.A crit-line monitor iv failed to recognize that the clic device was attached, and the device¿s red light was not lit.The usb cable continuity test passed.An internal inspection of the clic device revealed corrosion and thermal damage on the power supply board.Corrosion was on the board, jumper j108, ferrite rz100, and fuse f100.The board and fuse f100 both exhibited thermal damage.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed, and the cause was determined to be physical damage on components of the power supply board.The device was returned to returned goods following the evaluation.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported to customer service (cs) that a clic device would not verify.The biomed confirmed the defective device was previously installed and provided treatment data; this was not an out-of-box failure.A replacement device was issued to the facility.It was confirmed that the replacement device was received, and the defective device was returned to the manufacturer for evaluation.Upon physical evaluation of the returned device by the manufacturer, evidence of thermal damage was identified on the power supply board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRIT-LINE CLIP (CLIC) - USB
Common Device Name
ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key12760634
Report Number2937457-2021-02238
Device Sequence Number16927345
Product Code KOC
UDI-Device Identifier00840861101139
UDI-Public00840861101139
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K121599
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCL10041001
Device Catalogue NumberCL10041001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Initial Report FDA Received Date11/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-