MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 CHANNEL DRAIN,19 FR., ROUND, HUBLESS, FULL FLUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 CHANNEL DRAIN,19 FR., ROUND, HUBLESS, FULL FLUTED Back to Search Results
Model Number 072230
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the coronary artery bypass graft surgery (cabg) patient had channel drain placed in the right pleural space on (b)(6) 2021.On (b)(6) 2021, the physician discontinued the drain.During routine removal, the tube broke into two pieces, leaving a portion of the tube inside of the patient.The patient had surgery next day to remove the retained portion.The separation of the two pieces were jagged and uneven.It broke about 1 inch above the black dot that was located towards the bottom of the drain.Nurse reported that it was a routine removal with normal resistance, and it was easily removed until it broke.Patient had been doing well and was expected to continue but after the surgery to remove the retained channel drain, the patient had not done well.Physician indicated the additional anesthesia required for the surgery may have affected the patient condition.Patient was insulin dependent diabetic, on hemodialysis, history of amputation of lower extremity and other co-morbidities.Patient is currently in icu on ventilator.Per form received on (b)(6) 2021, stated that there was a surgical removal of retained tubing.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1samples were confirmed to exhibit the reported failure.The device had not met specifications.The product was used for patient treatment.Visual evaluation of the returned sample noted one opened (without original packaging), used channel drain tube.Visual inspection of the sample noted obvious visible defects breakage into two pieces on the return sample.This is considered a failure stating that "no cuts or tears are allowed on the surface of the drains.A dhr review is not required as the lot number is unknown.The investigation is concluded, and no additional action is required at this time.The device was returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: indications: wound drains are used to remove exudates from wound sites.Warnings: an effective closed suction drain system requires maintenance of the system to preserve patency.The drain must not be allowed to occlude nor the reservoir to completely fill; and reservoir suction must be maintained in order for the system to function properly.Verify that the system is functioning properly.If the system is not maintained properly, surgical complications, including hematomas, may result.In the event of occlusion of the drain, all wound drainage via the drain ceases.Careful attention to the drain will minimize the possibility of this problem.If occlusion does occur, the drain can be aspirated by connecting suction to the reservoir outlet or temporarily disconnecting the drain from the reservoir and applying suction directly to the drain.If an air-tight seal between the drain and the skin where the drain emerges is not achieved, the air leak must be rectified or the system must be converted to open drainage.An airtight seal between all system components (drain, adaptor and reservoir) is necessary for proper system function.Leaving the soft silicone elastomer drain implanted for any period of time so as to cause tissue ingrowth around the drain can interfere with easy removal and may effect the performance of the drain.The surgeon should monitor the patient¿s rate of wound healing.Evacuators should be used in cardio-thoracic surgery only after the lung is fully expanded and all air leaks have sealed.Drain perforations or channels must lie within the wound or cavity to be drained, otherwise inadequate drainage may result.To avoid the possibility of drain damage or breakage: avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Complications complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an air-tight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.In the event an air-tight seal is not achieved, the reservoir will rapidly fill with air from the leak; subsequent drainage to the reservoir will occur only if allowed by gravity and wound exudate forcing the flow.Entry into the reservoir is allowed only by displacement of air in the reservoir by wound exudate flow.In this displacement process, air reflux from the reservoir to the wound can occur and increase the likelihood of back-contamination across the anti-reflux valve.In the event of drain occlusion by fibrin, clots, or other particulate matter, all wound drainage via the drain ceases.If the reservoir is not emptied when it is full, equilibrium between the drain and reservoir at wound pressure will ultimately occur and drainage from the wound site will cease.When the reservoir and drain are at the same pressure and the reservoir is full of fluid, the likelihood of back-contamination across the anti-reflux valve is increased.When used to drain the pleural cavity in the presence of an air leak, drains must be attached to an appropriate pleural cavity drainage system to prevent tension pneumothorax.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the coronary artery bypass graft surgery (cabg) patient had channel drain placed in the right pleural space on 04oct2021.On 06oct2021, the physician discontinued the drain.During routine removal, the tube broke into two pieces, leaving a portion of the tube inside of the patient.The patient had surgery next day to remove the retained portion.The separation of the two pieces were jagged and uneven.It broke about 1 inch above the black dot that was located towards the bottom of the drain.Nurse reported that it was a routine removal with normal resistance, and it was easily removed until it broke.Patient had been doing well and was expected to continue but after the surgery to remove the retained channel drain, the patient had not done well.Physician indicated the additional anesthesia required for the surgery may have affected the patient condition.Patient was insulin dependent diabetic, on hemodialysis, history of amputation of lower extremity and other co-morbidities.Patient is currently in icu on ventilator.Per form received on 27oct2021, stated that there was a surgical removal of retained tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHANNEL DRAIN,19 FR., ROUND, HUBLESS, FULL FLUTED
Common Device Name
CHANNEL DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12761165
Report Number1018233-2021-06991
Device Sequence Number4217776
Product Code GBX
UDI-Device Identifier00801741049903
UDI-Public(01)00801741049903
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number072230
Device Catalogue Number072230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/14/2021
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date11/07/2021
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
Patient Weight75 KG
-
-