| Catalog Number |
046W1AN00641 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3003853072-2021-00118 through 3003853072-2021-00120.
|
| |
|
Event or Problem Description
|
|
It was reported the tips of two final screwdriver shafts broke intra-operatively while final tightening the blockers.The fragments were recovered and the procedure was completed using another final driver shaft without patient impacts.This is report three of three for this event.
|
| |
|
Event or Problem Description
|
|
It was reported the tips of two final screwdriver shafts broke intra-operatively while final tightening the blockers.The fragments were recovered and the procedure was completed using another final driver shaft without patient impacts.This is report three of three for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed fractured tips on 3 instinct java final screwdriver shafts.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to off-axis forces applied during use.Dhr review: per dhr review, the parts were likely conforming when they left zimmer biomet control.Device use: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|