Model Number 9004051
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that no wave and not responding to breaths, giving intermittent issues.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Malfunctioning co2 pump mechanism was unable to draw adequate airflow through the pneumatics.A good known co2 pump was swapped to bring the airflow to within range.However, the tamper seal was broken and the main board was dislodged from the back panel referring to user tampering and/or drop damage.Therefore, unable to determine what caused the co2 pump failed.Further dhr review is not relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.
 
Additional Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2021-10830.The report was submitted in error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI MONITOR
Common Device Name
CCK
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12763904
Report Number3012307300-2021-10830
Device Sequence Number19915036
Product Code CCK
UDI-Device Identifier10610586036873
UDI-Public10610586036873
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K970209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9004051
Device Catalogue Number9004051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2021
Supplement Date Received by Manufacturer12/13/2021
05/02/2023
Initial Report FDA Received Date11/07/2021
Supplement Report FDA Received Date01/11/2022
05/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-