• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TC3 RP TIBIAL INSERT S4,20.0; SIGMA REVISION IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TC3 RP TIBIAL INSERT S4,20.0; SIGMA REVISION IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 96-2355
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Sepsis (2067); Joint Contracture (4528); Unspecified Musculoskeletal problem (4535); Swelling/ Edema (4577)
Date of Event 10/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
(b)(6).Clinical notification received for revision due to infection.Tibial insert exchange only.Date of implant: (b)(6) 2012, date of revision: (b)(6) 2021, (left knee).
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TC3 RP TIBIAL INSERT S4,20.0
Common Device Name
SIGMA REVISION IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12770083
Report Number1818910-2021-24577
Device Sequence Number19612780
Product Code NJL
UDI-Device Identifier10603295242413
UDI-Public10603295242413
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number96-2355
Device Catalogue Number962355
Device Lot Number109602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2021
Supplement Date Received by Manufacturer12/09/2021
Initial Report FDA Received Date11/08/2021
Supplement Report FDA Received Date12/10/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MBT REVISION CEM TIB TRAY SZ 4; PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 4MM; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA DIS AUG 4MM,SZ4,LEFT; PFC*SIGMA TC3 FEM LEFT SZ4; PFC*SIGMA/OV/DOME PAT 3PEG,41; SIG FEM ADAP +2/-2 OFFSET BOLT; SIGMA FEM ADAPTER 5 DEGREE; TAPER CEM STEM 13X150MM; TC3 RP TIBIAL INSERT S4,20.0; UNIVERSAL FEM SLV DIS POR 40MM; UNIVERSAL STEM 75X16MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
Patient Weight116 KG
Patient EthnicityHispanic
-
-