| Model Number |
320-46-10 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
10/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer? pending evaluation: concomitant medical device(s): 320-01-46, (b)(4) - eq rev 46mm glenosphere; 320-10-05, (b)(4) - eq rev adapter plate tray +5; 320-15-05, (b)(4) - eq rev locking screw; 320-20-00, (b)(4) - eq rev torque screw kit.
|
| |
|
Event or Problem Description
|
|
As reported approximately 22 months postop the initial ltsa this (b)(6) y/o male patient reached down to pull up a kickstand on his motorcycle and felt a pop.He was revised to the following, downsizing from 46mm to 42mm.Patient was last known to be in stable condition following the event.Device will not return due to facility policy.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported may have been the result of either incomplete seating of the liner during implantation, bone impingement, patient-related conditions, post-traumatic event, or any combination of these possibilities, which led to humeral liner disassociation.However, this cannot be confirmed as the devices were not returned for evaluation and x-rays were not provided.
|
| |
|